Phone: (+33) 625 16 20 71 - Sunday to Friday contact@alliancemedicalfrance.fr

Olympus Piston d’Aspiration MAJ-207

Olympus

Piston d’Aspiration MAJ-207

Olympus MAJ-207 reusable suction valve designed for controlled suction during endoscopic procedures, providing reliable fluid and debris evacuation.

Olympus MAJ-207 – Suction Valve

The Olympus MAJ-207 is a manual suction valve designed for use during endoscopic procedures with compatible Olympus BF-series bronchoscopes. It provides a practical means of activating suction to remove fluids and residues that may interfere with visualization during bronchoscopic examinations.

Manual Suction Control

The MAJ-207 is operated by applying pressure to the valve button with the index finger, activating the suction mechanism integrated into the endoscope. This allows the operator to remove fluids or debris and maintain a clearer endoscopic field during the procedure.

Secure Installation

The suction valve is designed to attach securely to the endoscope's suction cylinder. Correct installation is essential to prevent gaps between the valve base and the instrument, which could otherwise result in the projection of biological material.

No lubricant is required for installation. Inappropriate lubrication may affect the seals and compromise the proper operation of the valve.

Installation and Alignment

The MAJ-207 is inserted into the suction cylinder by applying pressure while following the visual alignment mark provided on the endoscope body. Correct positioning ensures proper engagement of the valve with the suction mechanism.

Reuse and Inspection

The MAJ-207 is reusable for a maximum of six examinations. Beyond this limit, continued use may increase the risk of reduced performance and cross-contamination.

A thorough visual inspection should therefore be performed before each use. Any deformation, scratches, or abnormal resistance of the activation button should result in immediate replacement of the valve.

Compatibility and Reprocessing

The MAJ-207 is designed for reprocessing in accordance with applicable sterilization and disinfection procedures for compatible Olympus BF-series bronchoscopes, including the BF-30, BF-40, BF-160, BF-200, BF-240, and BF-260 series.

Its configuration is intended to provide secure integration with compatible Olympus endoscopic equipment used in hospital and specialized clinical environments.

The Olympus MAJ-207 provides a practical solution for suction control during compatible bronchoscopic procedures. Its manual operation, secure installation, and reusable configuration support effective handling and maintenance of the endoscopic field.

Quick delivery worldwide
All our equipment is rigorously tested to ensure its quality and compliance with safety standards.
Download brochure

Olympus MAJ-207 – Suction Valve

The Olympus MAJ-207 is a manual suction valve designed for use during endoscopic procedures with compatible Olympus BF-series bronchoscopes. It provides a practical means of activating suction to remove fluids and residues that may interfere with visualization during bronchoscopic examinations.

Manual Suction Control

The MAJ-207 is operated by applying pressure to the valve button with the index finger, activating the suction mechanism integrated into the endoscope. This allows the operator to remove fluids or debris and maintain a clearer endoscopic field during the procedure.

Secure Installation

The suction valve is designed to attach securely to the endoscope's suction cylinder. Correct installation is essential to prevent gaps between the valve base and the instrument, which could otherwise result in the projection of biological material.

No lubricant is required for installation. Inappropriate lubrication may affect the seals and compromise the proper operation of the valve.

Installation and Alignment

The MAJ-207 is inserted into the suction cylinder by applying pressure while following the visual alignment mark provided on the endoscope body. Correct positioning ensures proper engagement of the valve with the suction mechanism.

Reuse and Inspection

The MAJ-207 is reusable for a maximum of six examinations. Beyond this limit, continued use may increase the risk of reduced performance and cross-contamination.

A thorough visual inspection should therefore be performed before each use. Any deformation, scratches, or abnormal resistance of the activation button should result in immediate replacement of the valve.

Compatibility and Reprocessing

The MAJ-207 is designed for reprocessing in accordance with applicable sterilization and disinfection procedures for compatible Olympus BF-series bronchoscopes, including the BF-30, BF-40, BF-160, BF-200, BF-240, and BF-260 series.

Its configuration is intended to provide secure integration with compatible Olympus endoscopic equipment used in hospital and specialized clinical environments.

The Olympus MAJ-207 provides a practical solution for suction control during compatible bronchoscopic procedures. Its manual operation, secure installation, and reusable configuration support effective handling and maintenance of the endoscopic field.

  • Fluid and Residue Removal: Facilitates the removal of fluids and debris during endoscopic procedures.
  • Clear Field of View: Helps maintain clear visibility throughout the examination.
  • Integration with the Suction Cage: Designed to integrate with the endoscope suction cage for convenient operation.
  • Ergonomic Design: Provides practical handling with automatic piston return after use.
  • Limited Use: Designed for a maximum of 6 uses to comply with applicable safety requirements.
Utilisation :
Manuelle, avec activation à l’index
Compatibilité :
Séries Olympus BF-30, 40, 200, 240, 260 et 160
Nombre maximal d’utilisations :
6 examens
Lubrification requise :
Aucune
Fixation :
Enclenchement par pression dans la cage d’aspiration
Type :
Piston d’aspiration
Nettoyage :
Nettoyage et désinfection manuels ou automatiques selon les protocoles Olympus
  • MAJ-210 – Single-use biopsy valve
  • MD-495 – Reusable biopsy valve
  • MAJ-222 – Suction cleaning adapter
  • MA-651 – Mouthpiece
  • BW-15B – Channel cleaning brush
  • BW-7B – Channel cleaning brush
  • MH-507 – Channel-opening cleaning brush
  • BW-15SH – Suction connector cleaning brush
Scroll